616KD-A-12-FC-NIST-TC
Series: Series 616KD Differential Pressure Transmitter Brand: Dwyer Instruments, LLC
Price
$1,986.00
CAD
Stock
100+ In Stock
- Accuracy - 0.25% Full Scale
- Pressure Range - Pa: 0 to 750
- Tag - No Tagging
- Calibration - Factory Calibration
- Certificates - NIST Traceable Calibration Certificate
- Cover - Terminal Cover
- Output - 4-20 mA
Specifications
Brand
Accuracy
- 616KD: ± 2 % of Full Scale
- 616KD-A: ± 0.25 % of Full Scale
- 616KD-B: ± 1 % of Full Scale
Approvals & Certifications
- CE
Compensated Temperature
- -6.67° to 50° C (20° to 122° F)
Consumption
- 24 mA - Max.
Electrical Connectors
- Screw Typical Terminal Block
Environmental Protection
- IP20
- NEMA 1
Loop Resistance
- 0 to 1250 ohm, Max.
- 4-20 mA Output (DC), (AC)
- 5K ohm Min.
- Rmax = 50 (VpsDC -10) ohm
- Rmax = 50 (1.4 VpsAC -12) ohm
Media
- Air & Non-Combustible
- Compatible Gases
Mounting Position
- Vertical with Pressure Connections Pointing Down
Options
- Aluminum Tag
- Factory Calibration
- Terminal Cover
Output
- 4-20 mA or Option with Field Selectable 0-10, 0-5, 2-10, 1-5 Volts
Power Requirement
- 10 V
- 10-35 Vdc (2-Wire, 3-Wire)
- 12-26 Vac (4-Wire)
- 4-20 mA
- 5 V
Pressure Range
- bar: 0 to 0.05
- Inches WC: 0 to 40
- kPa: 0 to 5
- psi: 0 to 0.725
Pressure Type
- Differential
Process Connection Material
- Dual Size to Fit 1/8” & 3/16” (3 mm & 5 mm) I.D. Rubber
- Dual Size to Fit 1/8” & 3/16” (3 mm & 5 mm) I.D. Vinyl Tubing
Sensor Orientation
- Vertical with Pressure Connections Pointing Down
Span & Zero Adjustments
- Push Button
Stability
- ± 1 % of Full Scale Per Year
Temperature Range
- -17.8° to 60° C (0° to 140° F)
Tube Connection
- 1/8” Tube
- 3/16” Tube
Update Time
- 1200 seconds