The Loop Leader PD683/PD688 is a loop-powered, 1/8 DIN digital panel meter designed for demanding process control applications.
The PD683/PD688 loop-powered indicator is built in a shallow-depth case with an FM Type 4X front and features loop-powered backlighting and a wide -30° to 65° C operating temperature range.
The Loop Leader PD688/PD683 is an Intrinsically Safe and Non-incendive, loop-powered, 1/8 DIN digital panel meter designed for hazardous process control applications.
The PD688/PD683 loop-powered indicator is built in a shallow-depth case with an FM Type 4X front and features loop-powered backlighting and a wide -30 to 65° C operating temperature range.
Features
4-20 mA input
5-Digit LCD, 0.6" (15.2 mm) high
FM type 4X, IP65 front
Shallow depth case 3.2" behind panel
2 V drop (5.7 V with backlight)
Loop-powered backlight standard
Custom engineering units and bar graph
Linear, square root, or programmable exponent
Maximum and minimum display
Operating temperature range: -30° to 65° C (-22° to 149° F)
Input (Square Root & Programmable Exponent): 10 to 100 % of Full Scale
Input: ± 0.03 % of Span ± 1 Count
Approvals & Certifications
ATEX II 1 G, Ex ia IIC T4, IP65, Ta = -40° to 65° C (-40° to 149° F)
CE
CSA: Non-Incendive Suitable for use in Class I, Div. 2, Grade A, B, C, D, Class II, Div. 2, Groups F, G, Class III, Div. 2,
FM: Intrinsically Safe with Entity for use in Class I, Div. 1, Grade A, B, C, D, Class II, Div. 1, Groups E, F, G, Class III, Div. 1, Class I, Zone 0, Group IIC, Temp. Code = T4
Calibration
Recalibration is Recommended at Least Every 12 Months
Communication
HART
Display
0.60” (15.2 mm) High
5-Digit LCD (-99,999 to 99,999)
7-Segment
Automatic Lead Zero Blanking
Backlight: Bright Orange LED (Intensity Varies with Signal)
Bargraph: 20-Segment, 0-100 % Indication
Display Update Rate: 2.5 Per second
Engineering Units: 0.25” (6.4 mm) High
Engineering Units: 14-Segment
Input (Low-Flow Cut-Off): -99,999 to 99,999
Electrical Connectors
Screw Terminals Accept 12 to 22 AWG Wire
Enclosure Finish
Gray
Enclosure, Body Material
Plastic
Stainless Steel
Steel
Environmental Protection
IP65
IP66
NEMA 4X
Function
Input: Linear
Input: Programmable Exponent
Input: Square Root
Humidity
0 % to 90 % Relative Humidity, Non-Condensing
Input
4-20 mA
Input Overload: over Current Protection to 2 A, Max.
Mounting
1/8” DIN Panel
Operating Temperature
-30° to 65° C (-22° to 149° F)
Programming
Four Front Panel Buttons
Resistance
Equivalent Resistance: 100 ohms at 20 mA without Backlight
Equivalent Resistance: 285 ohms at 20 mA with Backlight
Span - Minimum
0.40 mA between Input 1 & Input 2
Storage Temperature
-40° to 85° C (-40° to 185° F)
Temperature Range
Allowable Temperature Range: -40° to 65° C (-40° to 149° F)
Thermal Drift
Input: 50 ppm/°C from -40° to 65° C (-40° to 149° F) Ambient