Ultra 4 Ultrasonic Controller

Brand: Pulsar Measurement

The Ultra 4 is a fully functional advanced controller for level, flow, volume, and pump control. MCERTS Class 1 certified, the Ultra 4 offers open channel flow measurement for stormwater overflows and event duration monitoring.

The multifunctional display offers the ability to view trend graphs on-screen and get a glance at how levels have changed over time. 

When combined with the dB3 with double sun shields, the Ultra 4 achieves MCERTS Class 1 certification, with 0.193% combined accuracy, meaning the system has undergone rigorous testing and has achieved a high standard in every single one. 

This means that wastewater processes across the world can achieve a common standard on their flow applications, providing real-time data on flow into storm tanks, influent flow, and event duration monitoring.

In addition to the Ultra 4's open channel flow features, the unit offers sophisticated level control, includes built-in volume calculations based on standard tank shapes, and advanced pump control functions.

The Ultra 4 can be programmed either by the built-in keypad (comes as standard on all wall and fascia units), via the SD card slot, or by PC via the RJ11 Serial Interface. All parameters are stored in non-volatile memory and are retained in the event of a power interruption. 

A second backup copy of all parameters can also be retained in the Ultra 4 memory, in case an alternative set of parameters needs to be stored.
Features
  • Built-in volume calculations from standard tank shapes or calibration curves
  • 32 GB Micro SD Card
  • Echo profiles and level trend graphs are shown on the built-in display
  • RS485 Modbus/Profibus option
  • Radar or Ultrasonic measurement technology
  • 4 onboard relays assignable by type
  • Wall or Fascia Mount 
  • mA Output
  • The ability to measure pump control, flow measurement, and volume measurement in one compact unit
  • Quick set up features allow the easy programming of parameters

Specifications

Weight

  • 700 g (1.5 lbs)

Dimensions

  • Wall Mount: 150 mm x 130 mm x 63.5 mm (5.9 in x 5.1 in x 2.5 in)
  • Fascia Mount: 160 mm x 180 mm x 63 mm (6.3 in x 7.07 in x 2.5 in)

Enclosure Material / Description

  • Polycarbonate, Flame Resistant to UL 94-V0

Cable Entry Detail

  • 3 x M20 Glands

Transducer Cable Extensions

  • 3-Core Screened

IP Protection

  • IP64
  • IP67/NEMA 4X

Operating Temperature

  • Electronics: -20° to +45° C (-4° to +113° F)

Accuracy

  • ± 2 mm (0.079 in) for dBR16 mmWAVE RADAR
  • 0.25 % of The Measured Range or 6 mm, whichever is greater

Resolution

  • 0.1 % of The Measured Range or 2 mm (0.079 in), whichever is greater

Analog Output

  • Isolated Floating Output (to 150 V) of 4-20 mA or 0-20 mA Into 1 k Ω, User-Programmable and Adjustable

Certifications & Approvals

  • CE
  • UL
  • MCERTS

show more