Accuracy
- Option: ± 0.25 % of the Measured Value ± 1.5 mm/sec, Max.
- Standard: ± 0.5 % of Measured Value ± 1 mm/sec, Max.
Applications
- Agriculture
- Air Conditioning Systems
- CIP- Systems
- Food & Beverage
- Heating Systems
- Irrigation Systems
- Steel Mills
- Ventilation Systems
- Water & Wastewater
- Water Distribution Network, Abstraction
Approvals & Certifications
- IEC 60068-2-3, IEC 61298-3, IEC 68-2-64
Approvals & Certifications
- CE - EU Directives
- NAMUR - NE 21, NE 43, NE 53
Cable Entry
- 1/2” PF
- Option: 1/2” NPT
- Standard: M20 x 1.5 (8 to 12 mm)
Cable Length
- Signal: 1,950 feet (600 metres) Depending on Electrical Conductivity & Flow Sensor
Cable Type
- Double-Shielded Cable - Transmission between the Flow Sensor & Signal Converter
Communication
- Baud Rate: 1,200, 2,400, 3,600, 4,800, 9,600, 19,200, 38,400, 57,600, 115,200 bit/sec
- HART
- Master/Slave
- Modbus RTU
- RS-485
Conductivity
- Electrical Conductivity - Water: ≥ 20 µS/cm
- Electrical Conductivity: ≥ 5 µS/cm (All Media Except for Water)
Control Functions Characteristics
Diagnostics
- Empty Pipe Detection
- Measuring Device
- Stabilization
Display
- 128 x 64 Pixels
- 2.32” x 1.22” (59 x 31 mm) - Ambient Temperature Below -25° C / -13° F May Affect the Readability of the Display
- LCD Display, Backlit White
Flow Rate
- 0.005 to 48,858 m³/hour
- 0.3 to 12 m/sec
Frequency
- Pulse Rate / Frequency: 0.01 to 10,000 Hz
Function
- Step - the Time Constant Corresponds to the Elapsed Time Until 67 % of the End Value Has Been Reached
Load Resistance
- Test Point: ≥ 250 ohms at HART
Operating Pressure
- psi: 0 to 72.52 (kPa: 0 to 500, bar: 0 to 5)
Output
- ± 10 µA, ± 100 ppm/°C (Typically: ± 30 ppm/°C)
- With & without HART: 10-21.5 mA, Q=100 %
- With HART: 4-20 mA, Q=0 %
- Without HART: 0-20 mA, Q=0 %
Power Supply
- 100-230 Vac, 50/60 Hz (-15 % / +10 %) Tolerance
- 24 Vdc (± 30 %) Tolerance
- 240 Vac - (5 % is Included in the Tolerance Range)
Principle of Operation
- Faraday's Law of Induction
Storage Temperature
- -40° to 70° C (-40° to 158° F)
Wiring
- Insulated Wire: 2 x 0.5 mm² Cu / AWG 20