Approvals & Certifications
Approvals & Certifications
- BS EN 61010, 60873, 61326
Battery Life
- Replacement Period: 3-Years
- Support Time (RTC): 1-Year Min. with Recorder UnPowered
Cable Length
- Ethernet Communications: 325 feet (100 metres) - Max.
Communication
- 10/100 BaseT Ethernet (IEEE 802.3)
- BOOTP
- DHCP
- FTP
- ICMP
- Modbus
- SNTP
- TCP/IP Protocols
- USB
Current
- AC Load Ratings, Max.
- Thermocouple Data (Bias Current): 0.05 NA
- Transmitter PSU: Max.
Display
- Operator Interface (Size & Resolution): 5.5” 1/4 VGA (320 x 240 Pixels)
- Operator Interface: TFT Color LCD with Cold Cathode Backlight, Fitted with Resistive, Analog, Touch-Panel
Electrical Connectors
- Ethernet Communications (Termination): RJ45
- Ethernet Communications: CAT5 Cable
- Input Board: Edge Connector/Terminal Block
Humidity
- Environmental Performance (Operation): 5 to 80 % Relative Humidity
- Environmental Performance (Storage): 5 to 90 % Relative Humidity
Input
- (Input Board) Channel to Ground, 1,500 Vac
- (Input Board) Open Circuit Detection, ± 57 NA, Max.
- DC Input Range: -150 to 150 mV
- DC Input Range: -20 to 20 V
- Input Board (Channel to Channel): 2,500 Vac
- Input Board (Over Voltage Protection): 50 V Peak (150 V with Attenuator)
- Input Board, Max.
Input Impedance
- Input Board: > 10 megaohms for 38 mV, 150 mV, 1 V Ranges
- Input Board: 65.3 kiloohms for 20 V Range
Insulation Resistance
- Input Board: > 10 megaohms at 500 Vdc
Isolation
- Input Board: 300 Vrms (Double, Basic Insulation)
- Transmitter PSU (Channel to Channel, Channel to Ground): 100 Vrms or DC (Double, Basic Insulation)
Linearization
- RTD Typical, Max.
- Thermocouple, Max.
Mounting
- Panel Mount: DIN 43700, ± 45°
Operating Temperature
- Environmental Performance: 0° to 50° C (32° to 122° F)
Output
- Transmitter PSU: 25 V Nominal
Power Requirement
- 240 Vac
- Maximum
- Poly-Carbonmonofluoride / Lithium Batteries
Power Supply
- 110-370 Vdc
- 90-264 Vac, 45 to 65 Hz
- Isolated Transmitter Power Supply: 110-240 V
- Isolated Transmitter Power Supply: 24 V
Resistance
- Input Board (Min. Break Resistance): 10 megaohms
- Resistance Input (Lead Resistance): 0 to 6 kiloohms
Resolution
- DC Input Range (Resistance): 5 to 148 milliohms
- DC Input Range: 1.4 to 720 µV
RTD Type
- Cu10
- Cu53
- JPt100
- Ni100
- Ni120
- Pt100
- Pt1000
- Pt100A
Sample Rate
- Update/Archive Rates (Input/Relay Output Sample Rate): 8 Hz
Sensor, Probe Type
- B Thermocouple
- C Thermocouple
- D Thermocouple
- E Thermocouple
- G2 Thermocouple
- J Thermocouple
- K Thermocouple
- L Thermocouple
- Ni/NiMo Thermocouple
- NiMo/NiCo Thermocouple
- N Thermocouple
- PlatineL Thermocouple
- R Thermocouple
- S Thermocouple
- T Thermocouple
- U Thermocouple
Shock
- Environmental Performance: BS EN 61010
Storage Temperature
- Environmental Performance: -20° to 60° C (-4° to 140° F)
Supply Voltage
- 100-230 Vac ± 15 %, 47 to 63 Hz
- Transmitter PSU: 110/120 Vac (100 mA)
Temperature Coefficient
- 0 to 5 kiloohms Resistance Range: 30 ppm
- -150 to 150 mV, -1 to 1 V Input Range: 25 ppm/°C
- -20 to 20 V Input Range: 388 ppm/°C
Temperature Range
- RTD Typical: -220° to 850° C (-364° to 1,562° F)
- Thermocouple: -270° to 2,495° C (-454° to 4,523° F)
Update Time
- Display, Trend Update: 1 second, Max.
Vibration
- Environmental Performance (10 to 150 Hz): BS EN 60873
Voltage
- AC Load Ratings, Max.
- The Rate Impulse Voltage for Equipment on Nominal 230 V Mains is 2,500 V