Programming Method
- Four CapTouch through-window buttons when cover is installed. The CapTouch buttons can be used with the cover removed. Free PC-based USB MeterView XL programming software.
Environmental
- Operating temperature range: -40 °C to 75 °C
- Storage temperature range: -55 °C to 85 °C
- Installation temperature range: -55 °C to 75 °C
Filter Bypass
- 0.0 to 99.9% of full scale. Input signal changes greater than bypass value are displayed immediately.
Connections
- Display terminals: accept 16 to 30 AWG wire
- Relay terminals: accept 12 to 26 AWG wire
- 4-20 mA output: accept 16 to 30 AWG wire
Enclosure Rating
- NEMA 4X, IP68 Explosion-proof
Accuracy
- ±0.02% of span ±1 count
- PD6908: Square root and programmable exponent: 10-100% FS
Input Overload
- Over current protection to 1 A maximum, Over voltage protection to 30 VDC max (between mA+ and mA-/BL-)
Temperature Drift
- 25 PPM/°C from -40 to 75°C ambient
EMC Emissions
- CFR 47 FCC Part 15 Subpart B Class A emissions requirements (USA)
- ICES-003 Information Technology emissions requirements (Canada)
- AS/NZS CISPR 11 Group 1 Class A ISM emissions requirements (Australia/New Zealand)
- EN 55011 Group 1 Class A ISM emissions requirements (EU)
- EN 61000-6-4 Emissions requirements for Heavy Industrial Environments - Generic
EMC Emissions and Immunity
- EN 61326-1
- EMC requirements for Electrical equipment for measurement, control, and laboratory use – industrial use
CSA Approvals
- Explosion-proof for use in: Class I, Division 1, Groups B, C and D
- Dust-ignition proof for use in: Class II/III, Division 1, Groups E, F and G; T6
- Flame-proof for use in: Zone 1, Ex d IIC T6 Ta = -55 to 75°C.
- Enclosure: Type 4X & IP66/IP68. Certificate number: 2325749
ATEX Approvals
- Intrinsically safe for use in: Ex ia IIC T4 Ga, Ex ia IIIC T200°C Da, Ta = -55 to 75°C
- Enclosure: Type 4X & IP66/IP68. Install per Control Drawing DW2636 Certificate number: CML 18ATEX2089X
- Explosion-proof for use in: ATEXII 2 G D, Ex db IIC T6 Gb, Ex tb IIIC T85°C Db IP68, Ta = -55 to 75°C, Certificate number: Sira 10ATEX1116X
IECEx Approvals
- Intrinsically safe for use in: Ex ia IIC T4 Ga, Ex ia IIIC T200°C Da, Ta = -55 to 75°C
- Enclosure: Type 4X & IP66/IP68. Install per Control Drawing DW2636, Certificate number: IECEx CML 18.0050X
- Explosion-proof for use in: Ex db IIC T6 Gb, Ex tb IIIC T85°C Db IP68, Ta = -55 to 75°C, Certificate number: IECEx SIR 10.0056X