Accuracy
- ± 10 µA, ± 100 ppm/°C (Typically: ± 30 ppm/°C)
- Optional: ± 0.2 % of Measured Value ± 1.5 mm/sec, Max.
- Standard: ± 0.3 % of Measured Value ± 1 mm/sec, Max.
Applications
- Chemical Industry
- Flow in Electrically Conductive Mediums with a Min. Conductivity of 5 µS/cm
- Food & Beverage Industry
- HVac Systems, Energy Management
- Machinery Industry
- Metal & Mining
- Water & Wastewater
Approvals & Certifications
- ATEX Ex
- CE - EU Directive
- CSA
- FM
- IEC 529, EN 60529, IEC 61158-2, IEC 68-2-27, IEC 68-2-64, IEC 664, IEC 364, IEC 536, IECEx
- NAMUR - NE 21, NE 43, NE 53
- NEPSI
Cable Entry
- 1/2” PF
- Option: 1/2” NPT
- Standard: M20 x 1.5 (8 to 12 mm)
Cable Length
- 1,950 feet (600 metres) - Depending on Electrical Conductivity & Flow Sensor
Cable Type
- Double-Shielded Cable - between the Flow Sensor & Signal Converter
Circuit Protection
- Reverse Polarity - PROFIBUS PA, FOUNDATION Fieldbus
Communication
- Baud Rate: 1,200, 2,400, 3,600, 4,800, 9,600, 19,200, 38,400, 57,600, 115,200 bit/sec - Modbus RTU, Master/Slave, RS-485
- FOUNDATION Fieldbus
- HART Communication Protocal
- Master/Slave
- Modbus
- PROFIBUS DP
- PROFIBUS PA - Class B, V3.02
- RS-485
Consumption
- Current: 10.5 mA (PROFIBUS PA, FOUNDATION Fieldbus)
- Power: 4 W (DC)
- Power: 7 VA (AC)
Control Functions Characteristics
Display
- 128 x 64 Pixels, LC Display
- Graphic: 2.32” x 1.22” (59 x 31 mm)
Enclosure, Body Material
- 316L Stainless Steel, 1.4401
- Aluminum
Environmental Protection
- IP66
- IP67
- IP69
- NEMA 4
- NEMA 4X
Flow Rate
- Volumetric: 0.3 to 12 (m/sec)
Frequency
- Modbus I/O: 0.25 to 1,000 Hz
- Pulse Rate / Frequency: 0.25 to 10,000 Hz
Load Resistance
- Test Point: ≥ 250 ohms at HART
Media
- Conductive Fluids
- Liquids
- Solids
- Water
Operating Pressure
- psi: 0 to 72.52 (kPa: 0 to 500, bar: 0 to 5)
Output
- With & without HART: 10-21.5 mA, Q = 100 %
- With HART: 4-20 mA, Q = 0 %
- Without HART: 0-20 mA, Q = 0 %
Power Supply
- 12 Vdc - 10 % is Included in the Tolerance Range
- 24 Vdc - (-55 % to 30 %), Only Available as Non-Ex Version
- 240 Vac + 5 % is Included in the Tolerance Range
- Optional: 24 Vac/dc (AC: -15 % to 10 %, DC: -25 % to 30 %), Non-Ex: Standard, Ex: Optional
- Standard: 100-230 Vac (-15 % to 10 %), 50/60 Hz, Non-Ex: Standard, Ex: Optional
Rotation
- 0° for Vertical Pipelines
- 45° for Horizontal Installation
- 90° Increments
Storage Temperature
- -40° to 70° C (-40° to 158° F)
Time Constant
- 0 to 100 seconds
- Set in Increments: 0.1 second
Wiring
- Insulated Wire: 2 x 0.5 mm² Cu / AWG 20